JOANLAB autoclavable single-channel micropipette · X series
Fully autoclavable single-channel unit, resistant to weak acids and alkalis, with nine color-coded ranges.
- The complete unit can be sterilized
- Resistant to weak acids and alkalis
- Nine ranges, from 0.2 µL to 10 mL
- Color-coded identification
Summary datasheet
| Models and ranges | X-2 (0.2-2 µL), X-10 (1-10 µL), X-20 (2-20 µL), X-50 (5-50 µL), X-100 (10-100 µL), X-200 (20-200 µL), X-1000 (100-1000 µL), X-5000 (0.5-5 mL), X-10000 (1-10 mL) |
| Stated systematic error | Stated by the manufacturer at three test volumes per model: the top of the range, its midpoint and its minimum. X-2: ≤±1.4% at 2 µL, ≤±2.5% at 1 µL and ≤±12% at 0.2 µL. X-10: ≤±1% at 10 µL, ≤±1.6% at 5 µL and ≤±7% at 1 µL. X-20: ≤±0.8% at 20 µL, ≤±1.2% at 10 µL and ≤±5% at 2 µL. X-50: ≤±0.8% at 50 µL, ≤±1.2% at 25 µL and ≤±4% at 5 µL. X-100: ≤±0.6% at 100 µL, ≤±0.8% at 50 µL and ≤±3% at 10 µL. X-200: ≤±0.6% at 200 µL, ≤±0.8% at 100 µL and ≤±3% at 20 µL. X-1000: ≤±0.6% at 1000 µL, ≤±0.8% at 500 µL and ≤±3% at 100 µL. X-5000: ≤±0.6% at 5 mL, ≤±0.8% at 2.5 mL and ≤±3% at 500 µL. X-10000: ≤±0.6% at 10 mL, ≤±0.8% at 5 mL and ≤±3% at 1 mL |
| Stated random error | Stated at the same three test volumes. X-2: ≤±0.7% at 2 µL, ≤±1.5% at 1 µL and ≤±6% at 0.2 µL. X-10: ≤±0.5% at 10 µL, ≤±1% at 5 µL and ≤±4% at 1 µL. X-20: ≤±0.5% at 20 µL, ≤±0.7% at 10 µL and ≤±2% at 2 µL. X-50: ≤±0.3% at 50 µL, ≤±0.5% at 25 µL and ≤±2% at 5 µL. X-100: ≤±0.2% at 100 µL, ≤±0.4% at 50 µL and ≤±1% at 10 µL. X-200: ≤±0.2% at 200 µL, ≤±0.3% at 100 µL and ≤±0.6% at 20 µL. X-1000: ≤±0.2% at 1000 µL, ≤±0.3% at 500 µL and ≤±0.6% at 100 µL. X-5000: ≤±0.2% at 5 mL, ≤±0.3% at 2.5 mL and ≤±0.6% at 500 µL. X-10000: ≤±0.2% at 10 mL, ≤±0.3% at 5 mL and ≤±0.6% at 1 mL |
| Compatible tips | 10, 200, 1000, 5000 and 10,000 µL depending on the model |
| Sterilization | Fully autoclavable, at 121 °C for 15 minutes. The manufacturer indicates about 10 cycles before recalibration |
| Setting increment | 0.002 µL (X-2), 0.01 (X-10), 0.02 (X-20), 0.05 (X-50), 0.1 (X-100), 0.2 (X-200), 1 (X-1000), 5 (X-5000) and 10 µL (X-10000) |
| Availability | Import to order / to be confirmed |
The only one declaring autoclave and chemical resistance together
Among the catalog's single-channel series, it is the only one carrying both statements at once: fully autoclavable and resistant to weak acids and alkalis. The Y series shares the first; chemical resistance, within pipetting, is otherwise declared only by the bottle-top dispensers.
It adds a four-digit window and an adjustment wheel, with color identification like the Y.
A finer step: increments down to 0.002 µL
The 2 µL model adjusts in 0.002 µL steps where the Y2.5 adjusts by 0.05 µL, and the pattern repeats up the ladder. At the midpoint of its table it also states less error than its Y counterpart: ≤±1.6% at 5 µL against ±2.4%.
The nine ranges cover 0.2 µL to 10 mL with the standard 10, 200 and 1000 µL, 5 and 10 mL tips.
It is worth looking at the bottom of the range, which the manufacturer states separately and where the cost shows: the X-2 goes from ≤±2.5% at 1 µL to ≤±12% at 0.2 µL, and the pattern repeats across the nine models. If the working volume sits near the minimum, the model below usually measures that same volume better.